Description
The Edan SE-1202 12-Channel EKG Machine (#SE-1202) includes an ECG Cable (4mm, banana connector, AHA), 100 resting tab electrodes, 10 Unomedical 10-Lead universal clips, Recording Paper (Z-fold, 210 mm x 295 mm x 100 P), a Power Cord (USA standard), a Rechargeable Lithium Battery (2200 mAH), a Fixed Bar for Battery (Necessary accessory for 2200 mAH battery), and two Fuses (1 A).
Warranty: 2 Years.
Powerful and sleek, the SE-1202 12-Channel EKG Machine provides a comprehensive set of cardiac diagnosis tools paired with cutting-edge to ensure the high-quality performance you've come to expect from Edan. It also features a glass-covered surface which supports easy cleaning and disinfection.
Features
Edan SE-1202 12-Channel EKG Machine Features
- Glasgow interpretation.
- Auto analyzing and diagnostic software (interpretation).
- Preview and printing for waveform and analysis.
- 10.1" LCD display with intuitive touch screen interface.
- USB data export.
- Wifi enabled.
- Built-in rechargeable Lithium battery.
- Built-in thermal printer.
Specifications
Edan SE-1202 12-Channel EKG Machine Specifications
- Dimensions: 400 mm x 360 mm x 90 mm.
- Weight: 14.33 lb (≤ 6.5 kg - excluding recorder paper and battery).
- Display: 10.1'', 1280 x 800 multicolor LCD Screen (the display can be flipped 25 degrees clockwise).
- Power Supply:
- Mains Supply:
- Operating Voltage: 100 V - 240 V~.
- Operating Frequency: 50 Hz/60 Hz.
- Input Current: 0.9-0.4 A.
- Internal Li-ion Battery Pack:
- Rated Voltage: 14.8 V.
- Rated Capacity: 2,500 mAh or 5,000 mAh.
- Necessary Charge Time: 3 hours or 6 hours.
- HR Recognition:
- HR Range: 30 BPM ~300 BPM.
- Accuracy: ±1 BPM.
- ECG Unit:
- Leads: 9 or 12 standard leads.
- A/D Converter: 24 bits.
- Time Constant: ≥ 5 s.
- Sampling Frequency: 64,000 Hz/channel.
- Input Impedance: ≥ 100MΩ (10Hz).
- Input Voltage Range: ≤ ±5 mVpp.
- CMRR: ≥140dB (AC on). ≥123dB (AC off).
- Acquisition Mode: 9 or 12 leads acquisition simultaneously.
- Resolution: 0.1192uV/LSB.
- Frequency Response: 0.01 Hz ~ 350 Hz (-3 dB).
- Gain: 1.25, 2.5, 5, 10, 20, 10/5 mm/mV, AGC, ±5%.
- Input Circuit Current: ≤ 0.01μA.
- Calibration Voltage: 1mV±2%.
- QTc Formulas: Bazzet, Fridericia, Framingham, Hodges, and QRS.
- Pacemaker:
- Amplitude: ±500μV to ±700 mV.
- Sampling Frequency: 80,000 Hz, Rhythm Lead.
- Width: 30μs to 2.0 ms.
- Filter:
- AC Filter: Off/50 Hz/60 Hz.
- EMG Filter: Off/25 Hz/35 Hz/45 Hz.
- DFT Filter: 0.01 Hz/0.05 Hz/0.32 Hz/0.67 Hz.
- LOWPASS Filter: 350 Hz/300 Hz/270 Hz/150 Hz/100 Hz/75 Hz.
- Environment Specifications:
- Temperature:
- Transport & Storage: -20º C (-4º F) ~ +55º C (+131º F).
- Working: +5º C (+41º F) ~ +40º C (+104º F).
- Atmospheric Pressure:
- Transport & Storage: 70 kPa ~106 kPa.
- Working: 70 kPa ~106 kPa.
- Relative Humidity:
- Transport & Storage: 15%~95% Non-Condensing.
- Working: 15%~95% Non-Condensing.