Description
The Maxtec MAX-1 Oxygen Cell is an FiO (Fraction of Inspired Oxygen) Sensor. It is compatible with a variety of systems and is primarily used with anesthesia machines, ventilators, O monitors, O analyzers; check the Configurations tab for specific compatibility information.
Warranty:
- Generic Equivalent: 18 Months - Free of Manufacturing Defects.
- Original: 1 Year Shelf Life.
Note: Can only be shipped within the U.S.
Specifications
Maxtec MAX-1 Oxygen Cell Specifications
- Electrical Interface: Gold plated slip rings.
#G0-010 (Generic Equivalent) specific:
- Response Time: < 12 seconds.
- Latex-free.
- Non-sterile.
- Weight: 0.2 lb.
- Certifications: FDA, CE, ISO10993-1, 5, 10:2003E, TUV, RoHS Compliant.
#R113P90 (Original) specific:
- Range of Measurement: 0 to 100%.
- Output Range: 14 to 20 mV.
- Zero Offset: ≤ 0.30 mV.
- Response Time: ≤ 17 seconds.
- Linearity: < 3% full scale.
- Expected Lifetime: > 500,000 %O hours.
- Operational Temperature Range: 10 to 40 °C.
- Storage Temperature Range: -20 to 50 °C.
- Humidity Range: 0 to 90% RH, non-condensing.
- Ambient Pressure Range: 700 to 1250 mbar.
- Interference: Per EN ISO 80601-2-55.
Configurations
Maxtec MAX-1 Oxygen Cell Configurations
- #G0-010: Generic Equivalent (Manufactured by Cables and Sensors):
- Compatible with Datex-Ohmeda models such as the 125, 205, 51, 510, 5100, 5125, 5140, 5200, 5250, 5250 RGM, 5400, 7200, 7800, 7810, Excel 210-7800, Excel 7800, Modulus 4700, Modulus 7800, Modulus II 7810, Oxicap, S5172802.
- OEM Part Number Cross References: Analytical Industries: PSR-11-915 / Biotest: S101 / CR: R-15 / Ceramatec: CAG-1 / City Technologies: MOX-6 / Datex-Ohmeda: 0237-2034-700 / Draeger - Air Shields: C-100 / Draeger: 6803290 / Envitec: OOM101 / Hamilton Medical: HM-10, HM-13 / Hudson RCI: 5556 / Integral Process: 65001 / MSA: 655264, 655263 / Mercury Medical: 10-103-17 / Nuova: E-15/0 / Philips: 15201A / Sensidyne: SV-14A / SensorTech: ST-10 / Sensoronics - Pacifitech: SS-10A, PT-10 / Teledyne: #C-41131 / VTI: VT1272, 101272, 100352.
- #R113P90: Original (Manufactured by Maxtec):
- Replacement For: Analytical Industries: PSR-11-915 / Datascope: Multinex Analyzer / GE Datex-Ohmeda: 0237-2034-700 Excel Modulus, 5250 RGM, 5150 Handheld Monitor / GE Ohmeda: 5120, 5100, 5250 RGM, 4700 Oxicap, 7800, 5125, 5150, Handheld Monitor, 0237-2034-700, Excel Modulus / Hewlett Packard: 15201A / Mercury Medical: 10-103-07 / Pacifitech: PT-10 / Teledyne: R15 #C-4113 / Teleflex (Hudson): 5556 / VTI: 100352, 101272, 100352.