Sale price$4,036.00
Earn 4036 points

Only 8 In Stock

Est. Delivery:

Shop with Confidence

Description

The Respironics Alice 5 Diagnostic Sleep System includes an amplifier/base station, head box/input box, and interconnecting cables.

Warranty: 90 Days - Parts and Labor.

Note: Compatible Respironics Alice Computer (#MFI-PHL-ALICE-LPTP or #MFI-PHL-ALICE-DESK) is sold separate.

The Alice 5 Diagnostic Sleep System sets new standards for sleep labs eager to customize powerful features today but interested in flexible architecture to accommodate tomorrow's growth.

Features

Respironics Alice 5 Diagnostic Sleep System Features

  • Fully integrated sleep laboratory system over a network cable or wireless.
  • Expanded channel capability to 55 total channels.
  • High quality ECG with six channels and Pulse Transit Time (PTT).
  • Real time impedance display.
  • Patient Communication/Intercom is accomplished over your network cable.

Specifications

Respironics Alice 5 Diagnostic Sleep System Specifications

  • Total Channels: 55.
  • Analog Inputs: 12 total, 10 in the base station, 2 in the headbox.
  • Neurological Channels:
    • Number of Channels: 26 (10 left, 9 right, 2 EOG, 5 EMG) 2 reference.
    • Input Impedance: 1.66 Mohm per Electrode 3.33 differential.
    • Bandwith: 0.32 Hz to 106 Hz.
    • Input Signal Range: +/- 3.3 mV.
    • Digital Resolution: 16 bits.
    • Sample Rate: 2000 Hz.
    • Max Storage Rate: 200 Hz.
    • Electrode Impedance Testing: 100 Hz square wave between patient electrodes.
    • Calibration Signal: Signal injection, 1 Hz, 98 µVpp.
    • Communication Interfaces: Signals sent from Patient Interface to Headbox through sensor cables. Data is sampled and sent to the Base Station where it is stored on a disk or sent through an ethernet connection to a Host PC.
    • During Analysis: 0.5 to 100 Hz low pass filter 50 or 60 Hz Notch filter Anti-aliasing filter.
  • Other Channel Types:
    • (2) Airflow: Thermal and Pressure Flow.
    • (1) Body Position.
    • (2) Actigraphy.
    • (6) ECG.
    • (1) Snore Mic.
  • Physical Characteristics:
    • Amplifier/Base Station:
      • 14" x 5" x 12.5" (35.6 cm x 12.7 cm x 31.7 cm).
      • Weight: 7.8 lb (3.5 kg).
    • Headbox:
      • 10" x 4" x 2" (25.4 cm x 10.2 cm x 5.1 cm).
      • Weight: 1.6 lb (0.72 kg).
  • System Power Requirements:
    • Power to Main Unit: 6.25 VDC, 3 to 6 Amps.
    • Power Supply: MW116 Power Supply 100 - 240 VAC, 50 - 60 Hz, 0.5 - 1.5 A.

Literature

Customer Reviews

Based on 1 review
100%
(1)
0%
(0)
0%
(0)
0%
(0)
0%
(0)
V
Vern G. (Georgia, United States)
"Refurbished" is a stretch

Alice 5 unit we received failed shortly after warranty expired, upon repair noticed the failed hard drive was over 10 years old and internal screws were missing. "Used and tested" would be a better description.

We're sorry to hear that your Alice 5 unit failed shortly after the warranty expired. Let us know if there is anything that we can assist with.