Description

The Smiths Medical CADD Plus 5400 (#CADD-5400) is an ambulatory pump that is designed for both human and veterinary use. Capable of intravenous and subcutaneous infusions, this top-of-the-line pump features flexible programming that can be put in either continuous or intermittent mode. In continuous mode the CADD Plus delivers medication at a constant rate; while in intermittent mode it delivers medication at regular pre-set intervals making it ideal for use in a wide variety of clinical settings or even the home.

Warranty: 6 Months.

Note: The following error codes are used to indicate a problem.

  • 33.
  • 99.
  • 47.
  • low battery.

Specifications

Smiths Medical CADD Plus 5400 Specifications

General

  • Medication Cassette Reservoir Capacity: 50 ml or 100 ml.
  • Pump Operating Temperature Range: 2° C to 40° C (35° F to 104° F).
  • Pump Storage Temperature Range: -40° C to 55° C (-40° F to 131° F).
  • Pump Humidity Range: 10% to 90% Relative Humidity.
  • Pump Resolution: 50 microliters per interval.
  • Size: 2.79 cm x 8.89 cm x 16.26 cm (1.1" x 3.5" x 6.4") including 50 ml Medication Cassette Reservoir.
  • Weight: 425 g (15 oz) including battery and empty 50 ml Medication Cassette Reservoir.
  • Pump Alarms: Low battery, battery depleted, pump in Stop mode, controller, microprocessor or motor fault, improper delivery, power up fault, low residual volume, RES VOL = 000, high delivery pressure (28 ± 12 psi).
  • Bolus Volume at Occlusion Alarm Pressure: < 0.25 ml.
  • Power Source: 9-volt alkaline battery (use Duracell Alkaline MN 1604 o Eveready Energizer Alkaline #522 for example).
  • Pump Timing Accuracy: ± 0.1%.
  • System Delivery Accuracy: ± 6%, nominal.
  • System Definition: System is defined as a CADD pump with an attached Medication Cassette Reservoir and CADD Extension Set with integral anti-siphon valve, or an attached CADD Administration Set with an integral o add-on anti-siphon valve.

Nominal

  • Continuous:
    • GIVEN Range: 000 - 999 ml in 1 ml increments rounded down to nearest ml.
    • RES VOL Alarm Range: 000 - 1,500 ml in 1 ml increments rounded up to the nearest ml. 000 ml for alarm off.
    • ML/HR Range: 0.00 - 75.0 ml/hr in 00.1 ml increments.
  • Intermittent:
    • ML Range: 0.00 - 150.0 ml, 00.1 ml increments.
    • PERIOD Range: 10 min - 12 hr, 10 min increments.
    • CYCLE Range: 20 min - 12 hr: 10 min, 10 min increments and 24 hours for a 24 hour cycle.
    • KVO ML/HR Range: 00.2 - 10.0 ml/hr, 10 min increments and 00.0 when not using KVO.
    • DELAY START Range: 10 min - 12 hr, 10 min increments and 0 hr: 00 min when not using DELAY START.
    • CYCLE GIVEN Range: 0 min - 12 hr: 10 min, in 1 min increments and 0 to 24 hours in 1 hour increments when CYCLE is programmed in 24 hour mode.
    • GIVEN Range: 000 - 999 ml in 1 ml increments rounded down to the nearest ml.
    • RES VOL Alarm Range: 000 - 1,500 ml in 1 ml increments rounded up to the nearest ml, 000 ml for alarm off.