- Weight: 0.2 lb
-
Learn more:
-
Purchasing Restriction: Medical license or prescription required
- MFR: SleepSense Ships from Manufacturer
- MPN: 1397
- Phone: (888) 757-7367
Frequently asked questions
Is this sensor exclusive to the Alice 3 system?
Yes, this sensor is designed specifically for integration with the Alice 3 monitoring system.
The key details confirming this are:
- It utilizes an Alice 3 specific connector, ensuring seamless plug-and-play functionality and reliable data acquisition.
- The configuration is fixed and optimized exclusively for use within the Alice 3 environment.
How does the sensor monitor respiration?
The SleepSense Effort Sensor is designed to capture and measure patient respiratory movements, which is crucial for polysomnography (PSG).
Here is how it functions:
- Technology: The device uses advanced piezo crystal technology to detect movements of the chest or abdomen.
- Conversion: It reliably converts these physical movements into a clear analog voltage waveform signal.
- Integration: This signal output provides clinicians with an accurate and reliable representation of the patient’s respiration waveforms for monitoring and diagnosis.
It is secured using a double buckle design and is specifically engineered to integrate with the Alice 3 monitoring system.
What is the benefit of the secure double buckle design?
The double buckle system, along with the incorporated Velcro tabs, ensures highly secure and stable positioning on the patient’s chest or abdomen.
This secure fastening mechanism is essential for:
- Maintaining consistent contact during sleep studies.
- Accurately converting physical respiratory movements into a clear, reliable analog voltage waveform signal.
Is a physician's order required for purchase?
Yes, this product is strictly intended for professional use only.
The SleepSense Alice 3 Double Buckle Piezo Crystal Effort Sensor is available for sale only to, or on the order of, a physician. This ensures its appropriate clinical application in professional sleep study settings.
What are the recommended filter settings?
The SleepSense Piezo Crystal Effort Sensor has the following recommended filter settings for optimal performance:
- Recommended High Pass Filter Setting: 0.1 Hz
- Recommended Low Pass Filter Setting: 70 Hz
These specifications are provided for effective integration with the Alice 3 monitoring system.
What are the operating and storage conditions for this sensor?
The SleepSense Piezo Crystal Effort Sensor is designed to operate and be stored within specific environmental parameters:
- Operating Temperature Range: 40 to 104 °F
- Storage Temperature Range: -4 to 140 °F
- Humidity (Operating and Storage): 5 to 95% (non-condensing)
Where on the body is this sensor typically placed?
The SleepSense Effort Sensor is designed for placement on a patient’s chest or abdomen.
It captures respiration movements, converting them into an analog voltage waveform signal to monitor respiratory effort during sleep studies. The sensor features a secure double buckle design and Velcro tabs to ensure stable, adjustable, and reliable positioning throughout the monitoring process.
What are the sensor's physical dimensions and materials?
The SleepSense effort sensor is built for durability and accurate placement. It features:
- Housing Material: A small, durable ABS box to protect the piezo crystal element.
- Sensor Dimensions: The housing measures 1” x 0.7” x 0.5”.
- Cable: The signal transmission uses white PVC insulation tinsel wire.
The sensor assembly also includes a secure fastening mechanism utilizing a double loop, buckles, and Velcro tabs for stable attachment to the chest or abdomen.
Is the sensor built for professional use?
Yes, the SleepSense Alice 3 Double Buckle Piezo Crystal Effort Sensor is specifically designed for professional use in clinical settings. It’s intended for healthcare professionals to monitor and interpret respiration waveforms in sleep studies and respiratory monitoring.
What components are included with the sensor?
The SleepSense Alice 3 Double Buckle Piezo Crystal Effort Sensor package comes with the following essential components for professional use:
- Piezo Crystal Sensor Unit: This includes the sensor housing and the secure double buckle/Velcro strap assembly designed for chest or abdominal placement.
- Alice 3 Specific Cable/Connector: This cable is specialized to ensure proper connectivity and integration with designated Alice 3 monitoring systems.
- User Manual/Literature: Referenced literature and the user manual are included, typically provided digitally.