Sale price$1,227.68
Earn 1228 points
Est. Delivery:

Shop with Confidence

Description

The B. Braun Outlook 400ES (#621-400ES) is a large volume pump with impressive and convenient drug library capabilities. Specially designed with the goal of eliminating errors with IV medication, this top-of-the-line pump utilizes innovative Configuration Management Application (CMA) software which allows for automatic drug library updates to be sent wirelessly to all network connected pumps.

Warranty: 6 Months.

Note: The following error codes are used to indicate a problem.

  • 63.
  • 75.
  • 86.
  • 123.

Features

B. Braun Outlook 400ES Features

  • Features an open architecture design allows for easier integration with external vendor applications.
  • Non-exclusive partnerships with pharmacy system vendors help speed up implementation.
  • Eliminates the need to manually update drug libraries.
  • Customized system solutions meet your infusion therapy needs.
  • Each pump utilizes a microprocessor to prevent channel confusion and catastrophic pump failures.
  • Extremely versatile - vertical stacking takes less space.
  • Non-numeric keypad with simple arrows helps prevent manual programming errors related to zero versus decimal entry.
  • Air trapping cassette catches up to 3 mLs of air reducing the nuisance air-in-line alarms.
  • Delivery mechanism makes for reliable and accurate operation.
  • Real-time display of dose and limit values during infusion enhances titration safety.
  • Quick and easy drug selection thanks to the active alphabetic matrix.
  • Single screen programming means you to not have to depend on memory.

Specifications

B. Braun Outlook 400ES Specifications

  • Type of Unit: Volumetric (Non Drop-Counting), Positive Pressure Displacement Reservoir.
  • Classification: Class 1-Internally Powered, CF type, IPX1 drip-proof, not suitable in the presence of flammable anesthetic and intended for continuous use in accordance with UL2601-1.
  • Bar Code Scanner: Class 2 in accordance with IEC 60825-1. Maximum Output = 1mW, Scan Rate=42 scans/sec.
  • Moisture Protection: IPX1; drip protected for horizontal usage.
  • EMI/EMC Compliance: Meets or exceeds EN60601-1-2:2001 for Class B, emissions and immunities standard.
  • Ground Impedance: Less than 0.2 ohms (tested per UL2601-1).
  • Electrical Shock Protection: Type CF shock protection in accordance with UL2601-1.
  • Power Requirements: 120 VAC, 2A, 60 Hz excluding accessories.
  • Grounding Resistance: Meets UL Standard 2601-1.
  • Leakage Current: Less than 300uAmps earth leakage (tested per UL 2601-1).
  • Plug: Hospital Grade (3 pin).
  • Dimensions: 12" W x 4.5" H x 8.75" D.
  • Weight: 11 lb.
  • Pole Size Range: 0.875” - 1.25”.
  • Operating Time of Battery with a Fully Charged Battery:
    • 7 hours @ 25 ml/hr (100ES/200ES).
    • 5 hours @ 25 ml/hr (300ES/400ES).
  • Delivery Rate Range:
    • 0.1 to 999.9 ml/hr, in 0.1 ml/hr increments.
    • Micro Mode: 0.1 to 99.9 ml/hr.
  • Volume to be Delivered Range: 0.1 to 9999.9 ml, in 0.1 ml increments.
  • Occlusion Limit Settings: 75, 100, 200, 300, 400, 500 and 750 mmHg. Standard default is 300 mmHg.
  • Keep Vein Open Rate: 3.0 ml/hr or selected rate, whichever is lower.
  • Volume Delivered Accuracy: ± 5%
  • Fluid Types: All standard IV fluids.
  • Alarms: Air-In-Line, Battery Very Low, Close Roller Clamp, Container Empty, Downstream Occlusion, Door Open, Hold Time Exceeded, KVO, Low Battery Warning, System Error.
  • Air-in-Line Alarm: ≥ 100 microliter bubble
  • Free-Flow Protection: Passive, set-based.
  • Memory: Permanent data retention after the pump is turned off unless the battery is depleted or disconnected, or unless the Data Retention Defaults have been otherwise selected in the Biomed Options.
  • Environmental Operating Limits: 65° F to 105° F (18° C to 40° C).
  • Maximum Bolus Volume (Microbore Tubing):
    • 0.047 ml at 399 ml/hr (75 mmHg).
    • 0.086 ml at 399 ml/hr (750 mmHg).
    • 0.053 ml at 500 ml/hr (400 mmHg).
    • 0.097 ml at 750 ml/hr (750 mmHg).
  • Maximum Time to Occlusion Alarm:
    • 24 min at 0.1 ml/hr (75 mmHg).
    • 33 sec at 399 ml/hr (75 mmHg).
    • 6 hours at 0.1 ml/hr (750 mmHg).
  • Maximum Volume Under Single Fault Condition: Incremental volume in case of single fault condition will not exceed 1.0 ml at a rate of 999.9 ml.

Literature

B. Braun Outlook 400ES Literature

B. Braun Outlook 400ES Operator's Manual

Customer Reviews

Be the first to write a review
0%
(0)
0%
(0)
0%
(0)
0%
(0)
0%
(0)